|
| |
| EVALUATION OF THE TEST
PROCEDURES FOR THE TESTING OF ANTIVIBRATION GLOVES PER ISO STANDARD 10819 |
| Reynolds, D.D.1 and Stein, J.K.1 1 Center for Mechanical & Environmental Systems
Technology (CMEST), University of Nevada, Las
Vegas, Nevada, USA |
INTRODUCTION
In the field of personal protective equipment, gloves
were being manufactured and marketed that claimed to significantly reduce the magnitude of
vibration transmitted from vibrating tools to the hand. Most of these claims proved to be
false. As a result, the International Organization for Standardization adopted ISO
Standard 10819 to define test procedures that must be used to measure the vibration
attenuation characteristics of gloves that were designed to reduce vibration into the
hand.(1) This standard also specified the vibration attenuation values that
must be achieved for a glove to be labeled as an "antivibration glove".
A project was undertaken at the Center for Mechanical
& Environmental Systems Technology (CMEST) at the University of Nevada, Las Vegas in
the USA to develop a test system that can be used to conduct glove vibration
transmissibility tests per the test procedures specified in ISO Standard 10819. The test
system that was developed during this project is described in a companion paper by the
authors of this paper.(3) The test protocol for measuring the vibration
transmissibility of gloves per the test procedures specified in ISO Standard 10819 were
examined and evaluated as part of this project. The results of this analysis are discussed
in this paper. |
ISO STANDARD 10819 TEST
PROCEDURES
ISO Standard 10819 specifies the test procedures that
must be used to measure the vibration transmissibility of gloves.(1) The
vibration transmissibility of a glove per ISO Standard 10819 is the ratio of the vibration
amplitude directed into the palm of the hand inside of a glove divided by the vibration
amplitude directed into the palm on the outside surface of the glove. The vibration
signals that are measured at the handle and into the palm are the overall acceleration
signals that are passed through an ISO weighting filter that is specified by ISO Standard
5349.(2) Figure 1 shows the ISO weighting filter. The vibration
transmissibility of a glove is a measure of the attenuation of vibration into the hand and
arm by means of a resilient or vibration-damping material placed in the glove. The lower
the vibration transmissibility, the more effective a glove is in reducing vibration energy
into the hand and arm.

Figure 1 ISO Weighting Filter per ISO
Standard 5349 and a Linear Weighting Filter |
ISO Standard 10819 specifies the amplitude of
vibration transmissibility that must be achieved for a glove to be classified as an
antivibration glove. The standard requires the overall vibration transmissibility of a
glove to be measured for mid frequencies (16-400 Hz) and for high frequencies (100-1,600
Hz). Figure 2 shows the third octave amplitudes of the vibration spectra for the mid and
high frequency test signals, respectively. Tables 1 and 2 shows the third octave band
amplitudes and required band tolerance values for the mid and high frequency test signals,
respectively. Vibration first corresponding to the mid frequency test spectrum and then to
the high frequency test spectrum are directed into the hand by means of a 40 mm diameter
handle attached to a vibration shaker. Sets of two measurements on each of three test
subjects for a total of six measurements are made for each frequency range. Three
different gloves, one for each test subject, are used for each test series. The six
individual transmissibility values for each of the mid and high frequency test signals are
averaged to obtain the respective average ISO Standard 10819 vibration transmissibility
values. The average mid-frequency transmissibility is designated TRM,
and the average high-frequency transmissibility is designated TRH. For a
glove to be classified as an antivibration glove:
TRM must be less than 1.0, and TRH
must be less than 0.6.
The resilient or vibration-damping material must be
placed in the palm and the full finger and thumb stalls of the glove.
\
Figure 2 ISO Standard 10819 Mid and High
Frequency Test Spectra |
|
|
Table 1
ISO Standard 10819 Mid Frequency Acceleration and Tolerance Values

|
|
|
Table 2 ISO Standard
10819 High Frequency Acceleration and Tolerance Values

|
|
| GLOVE VIBRATION TRANSMISSIBILITY TESTS |
Table 3 Legend for
Gloves that Were Tested

|
Thirteen
different gloves were tested. The type of vibration-damping material that was placed in
each glove is listed in Table 3.
Manual Method of Controlling Shaker Handle
Acceleration Values
The ISO Standard 10819 mid and high frequency test
spectra for this project were manually adjusted for the glove vibration transmissibility
tests. Three methods for adjusting the shaker handle acceleration spectra were examined:
Method 1: The third octave mid and high
frequency acceleration values of the shaker handle were adjusted with no one clasping the
handle. When the handle was clasped for a test, the acceleration amplitudes in the third
octave frequency bands below 160 Hz decreased as is shown in Figure 3. ISO Standard 10819
tests were conducted where no adjustments were made to compensate for these decreased
acceleration values. |

Figure 3 Typical Handle Mid
Frequency Acceleration Spectra (Spectrum-M) and High
Frequency Acceleration Spectra (Spectrum-H) for Methods 1, 2, and 3 (The
triangular symbols represent the lower and upper acceleration limits specified
by ISO Standard 10819.) |
Method 2: The acceleration
values of the shaker handle were adjusted per the procedures used in Method 1. The
decreases in third octave acceleration values were measured for each of the three test
subjects that were used for a specific test series. The corresponding average values were
calculated for each affected third octave frequency band. The third octave graphic
equalizer that was used to control the input spectrum to the shaker handle was then
adjusted to compensate for the "averaged" changes in acceleration values. Figure
3 shows the effectiveness of this method in controlling the ISO Standard 10819 mid and
high frequency acceleration values of the shaker handle. ISO Standard 10819 tests were
conducted with no further adjustments being made to the shaker handle vibration
amplitudes.
Method 3: The acceleration
values of the shaker handle were adjusted per the procedures used in Method 1. As each
test subject clasped the shaker handle for a test, the third octave graphic equalizer was
manually adjusted until the handle acceleration values for the test were within the third
octave amplitude limits specified by ISO Standard 10819. Method 3 was used for all of the
"official" ISO Standard 10819 vibration transmissibility tests that were
conducted at CMEST.
ISO Standard 10819 vibration transmissibility tests
were conducted using the ISO weighting filter as specified by the standard and a linear
weighting filter (Figure 1). Table 4 shows a comparison of the test results for Methods 1,
2, and 3. Table 4 indicates that the method of adjusting the shaker handle acceleration
values had little effect on the ISO and linear glove transmissibility values. |
|
Table 4 Test Results for
Gloves 1, 9, and 12 Using
Methods 1, 2, and 3

|
|

Figure 4 Filter Setup for
Measuring the ISO Weighted Acceleration Value |
Weighting Filter
Bandwidths
ISO Standard 10819 requires that the rms
acceleration values at the shaker handle and at the palm of the hand be measured using the
ISO weighting filter that is specified in ISO Standard 5349. Figure 4 shows a schematic of
the filter setup for measuring the ISO weighted acceleration amplitudes. Figure 1 shows
the overall results of the filter setup when the corner frequencies of the bandpass filter
are 6.3 Hz and 1,250 Hz.
|
Table 5
Bandpass Filter Corner Frequencies

|
|
Different laboratories
have interpreted the above requirements differently. Table 5 shows how the requirements of
ISO Standard 10819 have been interpreted by the Berufagenossenschaftliches Institut fur
Arbeitssickerheit (BIA), Delta Acoustic & Vibration (DELTA), and the Center for
Mechanical & Environmental Systems Technology (CMEST). Each of these laboratories have
selected different corner frequencies for the bandpass filter for the mid and high
frequency measurements. Table 6 shows a comparison of the vibration transmissibility test
results that were obtained at CMEST and the corresponding results that were obtained at
DELTA and BIA.
Table 6 indicates that the ISO Standard vibration
transmissibility results that were obtained at the above three laboratories agreed well
with each other. This good agreement occurred because the ISO mid and high frequency
spectra to the shaker handle was properly controlled at the three laboratories. Vibration
energy outside of the frequency bandpasses for the mid frequency spectrum (16-400 Hz) and
for the high frequency spectrum (100-1,600) were properly attenuated (at least -12
dB/octave at the low and high corner frequencies). When this is the case, any of the
corner frequencies given in Table 5 can be used for a bandpass filter that is placed
before an ISO weighting filter and yield similar results.
|
Table 6
Comparison of ISO Standard 10819 Glove Vibration Transmissibility Test Results

|
|
ISO Weighted Versus Linear Vibration
Transmissibility Values
ISO Standard 10819 requires that glove vibration
transmissibility values be obtained using the ISO weighted acceleration values at the
shaker handle and palm of the hand. One of the problems with this is that the ISO weighted
vibration transmissibility values tend to underestimate the vibration attenuation
characteristics of antivibration gloves. Figure 5 shows the effects of the ISO weighting
filter on the mid and high frequency input spectra for ISO Standard 10819. This figure
indicates that the ISO weighting biases the ISO weighted values of acceleration to the
third octave acceleration values that are closest to the lower frequency limits of the
respective input spectra. Thus, the attenuation effectiveness of a glove at the third
octave frequencies near the upper frequency limits of the respective input spectra is
significantly underestimated.
|
 Figure
5 Effects of the ISO Weighting Filter on the Mid and High
Frequency Vibration Input Spectra |
|
This problem can be addressed by measuring the
glove vibration transmissibility values, using a linear weighting filter (Figure 1). Table
7 shows the ISO and linear weighted vibration transmissibility values of the gloves that
were tested during this project. There was not a significant difference between the ISO
and linear transmissibility values for the mid frequency vibration input. However, the
linear transmissibility values were substantially less than their corresponding ISO
weighted values for the high frequency input.
|
Table 7 Summary
of ISO 10819 Vibration Transmissibility Test Results

|
|
Figure 6 shows the relations that exist
between the ISO and linear weighted vibration transmissibility values for the ISO 10819
mid and high frequency input spectra. The equation for the mid frequency input spectrum
is:

The correlation coefficient for this equation is 0.98
and the standard error of fit is 0.028. The equation for the high frequency input spectrum
is:

The correlation coefficient for this equation is 0.99
and the standard error of fit is 0.042. These two equations can be used to convert from
ISO weighted vibration transmissibility criteria values to their corresponding linear
criteria values. This yields the following:
TRM(Lin) < 1 and TRH(Lin) <
0.37.
|
|
 Figure 6 Relationship between ISO and Linear Weighted Vibration
Transmissibility Values |
|
OBSERVATIONS
ASSOCIATED WITH ISO STANDARD 10819 TESTS
Observations Associated with Test
Subjects
The test subjects had a significant effect on the
vibration transmissibility test results. Proper posture during a test was very important,
There were a few test subjects who had consistently high transmissibility values, while
there were a few who had consistently low transmissibility values. Proper training of the
test subjects was necessary to obtain reliable and repeatable test results.
Those test subjects who consistently had unusually
high transmissibility values could result in a glove not meeting the requirements of ISO
Standard 10819 to be classified as an antivibration glove.
It took training and great care to ensure that the
palm accelerometer adapter was properly placed between the palm of the hand and the glove
during a test.
When only three test subjects are used for the ISO
Standard 10819 tests, a test subject who consistently has unusually high vibration
transmissibility tests results can negatively bias the transmissibility tests for a glove.
Increasing the number of test subjects to four or five can minimize this problem.
Observations Associated with the Test
Procedures
It was very difficult to generate the mid and high
frequency vibration input spectra within the amplitude band limits that are specified by
ISO Standard 10819 without the use of a vibration feedback controller.
Allowing the mid and high frequency input spectra to
deviate from the amplitude band limits specified by ISO Standard 10819 did not have a
significant effect on the measurement of the mid and high frequency vibration
transmissibility values.
Vibration energy outside of the lower and upper
frequency band limits of the mid and high frequency test spectra must be sufficiently
attenuated to prevent this energy from negatively biasing the mid and high frequency test
results. This is particularly important for the high frequency test spectra.
If a feedback vibration controller is used to
generate the vibration test signals, the controller output must be checked to ensure that
the controller significantly attenuates the controller output signal outside of the lower
and upper frequency limits of the test signal.
When a vibration feedback controller is not used to
generate the vibration test signals, a bandbass filter must be placed after the signal
generator to attenuate the vibration signal outside of the test frequency bandpass. This
filter must have a minimum attenuation of -12 dB/octave at the lower and upper frequency
limits of the test signal.
When the ISO Standard 10819 test signals are properly
processed before the signals enter the power amplifier for the electromechanical shaker,
it does not make any difference whether or not the output signals from the accelerometers
on the shaker handle and on the palm accelerometer adapter are directed through bandpass
filters before they are directed through ISO weighting filters. If the test signals are
not properly processed before they enter the shaker power amplifier, the use of bandpass
filters before the ISO weighting filters and the selection of the corner frequencies for
these filters can have a significant effect on the ISO Standard 10819 vibration
transmissibility test results.
ISO weighting of the acceleration signals for the ISO
Standard 10819 tests biases the measured glove vibration transmissibility values to the
lower frequency limits of the mid and high frequency test signals. This often causes the
ISO weighted vibration transmissibility values to underestimate the effectiveness of a
glove in attenuating vibration to the hand. This is particularly true for the high
frequency vibration transmissibility tests. This problem can be resolved by using linear
weighting filters instead of ISO weighting filters (Figure 1).
Shaker handle resonance frequencies within the test
frequency band limits of the mid and high frequency test signals can negatively bias the
ISO Standard 10819 test results. This is particularly true for the high frequency
vibration transmissibility tests. A shaker handle should be tested for resonance
frequencies before it is used for ISO Standard 10819 tests. If resonance frequencies are
found to exist within the test signal frequency limits, the handle must be redesigned to
eliminate the resonance frequencies.
Sharp edges on the vibration-damping material side of
the palm accelerometer adapter can cause the adapter to "dig" into the
vibration-damping material. This can negatively effect the ISO Standard 10819 vibration
transmissibility test results. This problem can be resolved by rounding the edges of the
accelerometer adapter.
|
CONCLUSIONS
Use only trained test subjects for ISO Standard 10819
vibration transmissibility tests.
Test subjects who consistently yield unusually low or
high vibration transmissibility values should not be used as test subjects.
Consideration should be given to increasing the
number of test subjects to four or five.
The allowable amplitude band limits for the mid and
high frequency test spectra should be increase to +2 for all third octave band
frequencies.
Care must be taken to ensure that the vibration
energy into the shaker power amplifier outside of the of mid and high frequency test
spectra lower and upper frequency limits is significantly attenuated.
The ISO weighting of the acceleration signals for the
ISO Standard 10819 vibration transmissibility tests causes the ISO Standard 10819
vibration transmissibility values to understate the effectiveness of a glove in
attenuating vibration to the hand.
Linear filters should be used instead of the ISO
weighting filters should be used on the acceleration signals before they are processed to
obtained the ISO Standard 10819 vibration transmissibility values.
When linear filters are used on the acceleration
signals to obtain the ISO 10819 vibration transmissibility values, the criteria levels
should be changed to TRM(Lin) < 1.0 and TRH(Lin)
< 0.37.
The shaker handle must not have any resonance
frequencies in the frequency bandpass of the mid and high frequency test signals.
The sharp edges on the bottom side of the palm
acceleration adapter must be rounded to prevent the adapter from "digging" into
the glove vibration-damping material during an ISO 10819 test.
|
REFERENCES
ISO Standard 10819: Mechanical vibration and shock
- Hand-arm vibration - Method for t he measurement and evaluation of the vibration
transmissibility of gloves at the palm of the hand, 1996, International Organization
for Standardization, Geneva, Switzerland.
ISO Standard 5349: Mechanical vibration -
Guidelines for the measurement and the assessment of human exposure to hand-transmitted
vibration, 1986, International Organization for Standardization, Geneva, Switzerland.
Reynolds, D.D. and Stein, J.K., Design and
Evaluation of an Inexpensive Test Fixture for Conducting Glove Vibration Tests per ISO
Standard 10819, 8th International Conference on Hand-Arm Vibration, Umea,
Sweden, June 9-12, 1998.
|
 Paper presented at:
8th International Conference on Hand-Arm Vibration
Umea, Sweden
June 9-12,1998 |
|